Calibration Curve Calculator

Enter your standards (concentration and signal) and the signals of your unknown samples. Get the calibration line, R², the concentration of each unknown, and the LOD and LOQ.

Standards (concentration, signal)

Standards: concentration then signal. Unknowns: one signal per line.

Updates as you type. Your data stays in this browser.
Results
Calibration linesignal = 0.04989 × conc + 0.002071
Sensitivity (slope)0.04989signal per unit concentration
Intercept0.002071should be close to the blank reading
R²0.9998
Residual SD (s)0.00336scatter of the standards about the line
LOD (3.3 s / m)0.222limit of detection
LOQ (10 s / m)0.673limit of quantification
Working range0 to 127 standards

Unknown samples

SignalConcentrationCheck
0.152.965in range
0.356.974in range
0.5510.98in range
0.713.99outside the standards (extrapolated)
Chart
0246810120.00.10.20.30.40.50.60.7Concentration (mg/L)AbsorbanceAbsorbancey = 0.04989x + 0.002071, R2 = 0.9998

How to use the Calibration Curve Calculator

  1. Paste the standards: concentration in the first column and the instrument signal (for example absorbance) in the second.
  2. Paste the signals of your unknown samples, one per line, in the second box.
  3. Read the calibration line, R², LOD and LOQ, and the concentration found for each unknown.
  4. Check the flags: a result outside the standards, or below the LOQ, should not be reported as it stands.

How the calculator works

It fits a least-squares line to the standards, signal = m × concentration + b, then inverts it for each unknown: concentration = (signal − b) / m. It also calculates the residual standard deviation s of the standards about the line, which sets the detection limits: LOD = 3.3 × s / m and LOQ = 10 × s / m. Those are the widely used ICH convention; if your method or standard specifies a different one, use the residual SD shown and apply it yourself.

Each unknown gets a check. If its signal is above the highest standard or below the lowest, the result is extrapolated and flagged, because the calibration was never measured there. If the concentration is below the LOQ it is flagged as too low to quantify reliably.

Worked example: seven standards, four unknowns

The sample standards run from 0 to 12 mg/L with absorbances from 0.002 to 0.602. The calculator gives signal = 0.04989 × conc + 0.00207 with R² = 0.9998 and a residual SD of 0.0034, which gives an LOD of 0.22 mg/L and an LOQ of 0.67 mg/L.

The four unknown signals 0.150, 0.350, 0.550 and 0.700 correspond to 2.96, 6.97, 10.98 and 13.99 mg/L. The last one is flagged: 0.700 is higher than the highest standard (0.602), so 13.99 mg/L is an extrapolation. The right response is to dilute that sample so it reads inside the range and measure it again.

SignalConcentration (mg/L)Check
0.1502.96In range
0.3506.97In range
0.55010.98In range
0.70013.99Outside the standards

Tips for a reliable calibration

Use at least five standards spread evenly over the range your samples will fall in, plus a blank, and prepare them independently rather than as a serial dilution. Check the residuals for a curve: R² of 0.999 can still hide a bend near the top of the range. Run a standard of known concentration alongside the unknowns as a check, and make a fresh calibration whenever the instrument, reagent or day changes. The full method, including how to judge the fit, is in the calibration curve guide.

To present the curve as a figure with error bars from replicate readings, use the scatter plot maker, which exports SVG, PDF and 600 DPI PNG.

Frequently asked questions

How do I calculate concentration from a calibration curve?
Rearrange the line signal = m × concentration + b to concentration = (signal − b) / m, and put in the sample’s signal. This calculator does it for every unknown you enter.
What are LOD and LOQ?
The limit of detection is the lowest concentration distinguishable from the blank, often 3.3 s / m. The limit of quantification is the lowest that can be measured with acceptable precision, often 10 s / m.
Why is my result flagged as outside the standards?
Its signal is above the top or below the bottom standard, so the line is being extended beyond measured data. Dilute or concentrate the sample and re-measure, or add a standard there.
Should the line pass through zero?
Not by force. The intercept should be close to the blank reading, and a large intercept is a sign of a problem.
Is my data private?
No. The calculation runs in your browser, so the numbers you enter are never sent to a server.